Physical Specifications
|
|
Dimension and Weight
|
Dimension
|
230mm(W)*90mm(H)*280mm(L)
|
|
Weight
|
standard configuration 2 kg(with battery)
|
|
Operation Environment
|
Power Requirements
|
AC 100-250V, 50/60Hz,
|
|
Power
|
20VA
|
|
Temperature
|
5-40℃
|
|
Humidity
|
15-93% non-condensing
|
|
Safety
|
Meet the requirement of IEC60601 series
|
|
Performance Specification
|
|
Display
|
LCD
|
|
Display mode
|
1 Plethysmogram waveform
|
|
Interface
|
Parameter cable interface
AC power input socket
|
|
Battery
|
Lithium battery inside, 3 hours for charging, 6 hours for continuous working
|
|
Alarm Indicator
|
Alarm indicator light, power indicator light pulse tone alarm sound button tones
|
|
Alarm
|
User-adjustable high and low limits
3-level audible and visual alarm
|
|
SpO2
|
|
Measurement range:
|
0%-100%
|
|
Resolution:
|
1%
|
|
Accuracy:
|
±2% (70-100%, Adult/Pediatric, non-motion )
±3% (70-100%, Neonate, non-motion )
±3% (70-100%, Adult/Pediatric/Neonate, motion )
0-69% unspecified
|
|
Alarm range:
|
Alarm high limit range 31-100%
Alarm low limit range 20-99%
|
|
Refreshing rate
|
0.5s
|
|
EtCO2 (Side stream CO2)
|
|
Measurement range
|
0-99mmHg
|
|
Accuracy
|
±2mmHg(0-40mmHg)
±5% of reading(41-76mmHg)
±10% of reading (77-99mmHg)
|
|
Resolution
|
1mmHg
|
|
Sampling rate
|
70m1/min 100m1/min
|
|
Sampling rate accuracy
|
0.15 or 15m1/min
|
|
Start-up time
|
30 Sec. the module enters the warming up status after the start up.
1 minute later, it enters the ready-to-measure status.
|
|
Respiration rate
|
3-150 rpm
|
|
Respiration rate accuracy
|
±2BrPM (0-70BrPM)
±5BrPM (>70BrPM)
|
|
Response time
|
When measured with a neonatal water trap and a 2.5 m-long neonatal sampling line:
< 3.5 s @ 100 ml/min
<4 s @ 70 ml/min
|
|
|
When measured with an adult water trap and a 2.5 m-long adult sampling line:
<5.5 s @ 100 ml/min
<7 s @ 70 ml/min
|
|
Delay time
|
When measured with a neonatal water trap and a 2.5m-long neonatal sampling line:
<3 s @ 100 ml/min
<3.5 s @ 70 ml/min
|
|
|
When measured with an adult water trap and a 2.5m-long adult sampling line:
<5 s @ 100 ml/min
<6.5 s @ 70 ml/min
|